Sr850 performance test record, Self tests, Dc offset – SRS Labs SR850 User Manual

Page 243: Common mode rejection, Amplitude accuracy and flatness

Advertising
background image

SR850 Performance Test Record

Serial Number

Tested By

Firmware Revision

Date

Equipment Used

1. Self Tests

Test

Pass

Fail

Power On Tests

____

____

Keypad

____

____

Knob

____

____

Main Memory

____

____

Video Memory

____

____

2. DC Offset

Input Coupling

Reading

Upper Limit

AC

_______

0.500 mV

DC

_______

0.500 mV

3. Common Mode Rejection

Frequency

Reading

Upper Limit

100 Hz

_______

30 µV

4. Amplitude Accuracy and Flatness

Sensitivity

Calibrator Ampl.

Lower Limit

Reading

Upper Limit

1 V

1.0000 Vrms

0.9900 V

_______

1.0100 V

200 mV

200.00 mVrms

198.00 mV

_______

202.00 mV

100 mV

100.000 mVrms

99.00 mV

_______

101.00 mV

20 mV

20.000 mVrms

19.800 mV

_______

20.200 mV

10 mV

10.000 mVrms

9.900 mV

_______

10.100 mV

Sensitivity

Frequency

Lower Limit

Reading

Upper Limit

200 mV

24 kHz

198.00 mV

_______

202.00 mV

200 mV

48 kHz

198.00 mV

_______

202.00 mV

200 mV

72 kHz

198.00 mV

_______

202.00 mV

200 mV

96 kHz

198.00 mV

_______

202.00 mV

7-25

Page 1 of 4