14 performance tests – SRS Labs SR850 User Manual

Page 232

Advertising
background image

7-14

Performance Tests

3) For each of the amplitudes listed below, perform steps 3a through 3c.

AC Calibrator Amplitudes

R Output Expand

1.0000 Vrms

1

100.00 mVrms

10

10.000 mVrms

100

4.000 mVrms

250

a) Set the AC calibrator to the amplitude in the table.

b) Press the keys in the following sequence:

<Expand>

Highlight the R expand.

Enter the R Output Expand from the table.

c) Wait for the R reading to stabilize. Record the value of R.

4) This completes the amplitude linearity test. Enter the results of this test in the test record at the end of

this section.