Chapter 2

Setup and Configuration

© National Instruments Corporation

2-11

NI CVS-1450 Series User Manual

Figure 2-4. Wiring a Third-Party Power Supply to the 4-Position Power Connector

Caution

Do not connect the CVS-1450 device main power to a source other than

24 VDC ±10%. Do not connect the CVS-1450 device isolated power to a source less than
5 VDC or greater than 30 VDC. Doing so could damage the CVS-1450 device.

The CVS-1450 device ships with a 4-position power connector that plugs
directly into the power input connector on the CVS-1450 device. To wire
power to the 4-position connector, complete the following steps:

1.

Wire the voltage output of the 24 VDC ±10% power supply to the main
voltage input, labeled V, on the 4-position connector.

2.

Wire the common-mode signal (ground) output of the power supply
to the common-mode signal input, labeled C, on the 4-position
connector.

If you are using a separate power supply for the CVS-1450 device isolated
outputs, connect the voltage output on the power supply to the isolated
power (Viso) on the 4-position connector. Connect the common-mode
signal (ground) on the power supply to the isolated common-mode signal
(Ciso) on the connector.

Note

If you do not require a separate power supply for the CVS-1450 device isolated

outputs, you can daisy-chain the V to the Viso and the C to the Ciso on the connector.

For information about grounding the CVS-1450 device chassis to earth
ground, refer to the

Earth Ground Connection

section of Chapter 3,

LEDs,

DIP Switches, and Connectors

.

1

4-Position Power Connector

2

Ferrule

3

Power Supply Wires

1

2

3