Chapter 3

Connecting Signals

SCB-68 Shielded Connector Block User Manual

3-2

ni.com

Figure 3-1. Summary of AI Connections

+

+

+

V

1

ACH

AISENSE

AIGND

+

+

V

1

ACH

AIGND

+

+

V

1

ACH(+)

ACH(–)

AIGND

+

+

V

1

ACH(+)

ACH(–)

AIGND

R

Refer to the

Using Bias Resistors

section for information on bias resistors.

Signal Source Type

Floating Signal Source

(Not Connected to Building Ground)

Grounded Signal Source

Examples:
• Ungrounded thermocouples
• Signal conditioning with

Isolated outputs

• Battery devices

Examples:
• Plug-in instruments with

nonisolated outputs

Input

Differential

(DIFF)

Single-Ended —

Ground

Referenced

(RSE)

Single-Ended —

Nonreferenced

(NRSE)

+

+

+

Common-

Mode

Voltage

Common-

Mode

Voltage

Common-

Mode

Voltage

+

Common-

Mode

Voltage

+

Common-

Mode

Voltage

Common-

Mode

Voltage

Refer to the

Using Bias Resistors

section for information on bias resistors.

+

+

V

1

ACH

+ V

g

AIGND

Ground-loop losses, V

g

, are added to

measured signal.

NOT RECOMMENDED

+

+

V

1

ACH

AISENSE

AIGND

R