Accuracy and resolution considerations, Open thermocouple detection – National Instruments SCB-68 User Manual

Page 50

Advertising
background image

Chapter 5

Adding Components for Special Functions

© National Instruments Corporation

5-5

SCB-68 Shielded Connector Block User Manual

Accuracy and Resolution Considerations

When you measure voltage to subsequently measure current, take the
following steps to maximize measurement accuracy:

1.

Refer to the accuracy tables in Appendix A, Specifications, of the DAQ
device user manual at

ni.com/manuals

.

2.

Use Equation 5-1 to determine the code width, which is the smallest
signal change that a system can detect.

3.

Divide code width by the resistor value to determine the minimum
current value you can measure.

(5-1)

In Equation 5-1,

range

defines the values between and including the

minimum and maximum voltages that the ADC can digitize. For example,
the range is 20 when you measure a signal between –10 to 10 V. Gain,
which is determined by the input limits of the application, is a value you
apply to amplify or attenuate the signal.

Gain is expressed in decibels and is defined as:

(5-2)

Resolution

, or the smallest signal increment that can be detected by a

measurement system, is either 12 or 16 bits, depending on the DAQ device.

Open Thermocouple Detection

As an option, you can build open thermocouple detection circuitry by
connecting a high-value resistor between the positive input and +5V.
A resistor of a few M

Ω or more is sufficient, but a high-value resistor

allows you to detect an open or defective thermocouple. If the
thermocouple opens, the voltage measured across the input terminals rises
to +5 V, a value much larger than any legitimate thermocouple voltage.
You can create a bias current return path by using a 100 k

Ω resistor

between the negative input and AIGND.

Code Width

Range

Gain

2

Resolution

×

-------------------------------------------

=

Gain

20 Log f

( )

=

Advertising