counting after receiving a low-to-high edge. The time lapse since receiving the edge equals the
counter value difference (loaded value minus read value) multiplied by the CLK period.

To perform frequency measurement, program a counter to be level gated and count the number
of falling edges in a signal applied to a CLK input. The gate signal applied to the counter GATE
input is of known duration. In this case, you program the counter to count falling edges at the
CLK input while the gate is applied. The frequency of the input signal then equals the count
value divided by the gate period. Figure 3-16 shows the connections for a frequency
measurement application. You can also use a second counter to generate the gate signal in this
application. In this case, program the second counter for a one-shot mode. This scheme needs an
external inverter to make the output pulse of the second counter active high.

+5 V

I/O Connector

CLK

GATE

OUT

DGND

Lab-PC Board

Gate

Source

Signal

Source

Counter

4.7 k

13

Figure 3-16. Frequency Measurement Application

The GATE, CLK, and OUT signals for Counters B1 and B2 are available at the I/O connector.
In addition, the GATE and CLK pins are pulled up to +5 V through a 4.7 k

resistor.

Figure 3-17 shows the timing requirements for the GATE and CLK input signals and the timing
specifications for the OUT output signals of the 8253.