Accuracy information – National Instruments PCI-6023E User Manual

Page 84

Advertising
background image

Appendix A

Specifications

PCI-6023E/6024E/6025E User Manual

A-2

©

National Instruments Corporation

Overvoltage protection

FIFO buffer size......................................512 S

Data transfers ..........................................DMA, interrupts,

programmed I/O

DMA modes ...........................................Scatter-gather

(Single transfer, demand transfer)

Configuration memory size ....................512 words

Accuracy Information

PCI-6025E Accuracy Information

Powered On

Powered Off

ACH<0..15>

± 42

± 35

AISENSE

± 40

± 25

Nominal Range (V)

Absolute Accuracy

Relative Accuracy

% of Reading

Offset

Noise + Quantization

(mV)

Temp

Drift

Resolution (mV)

Positive

FS

Negative

FS

24 Hours

90 Days

1 Year

(mV)

Single Pt.

Averaged

(%/

°

C)

Theoretical

Averaged

10

–10

0.0722

0.0742

0.0764

±

6.385

± 3.906

± 0.975

0.0010

4.883

1.284

5

–5

0.0272

0.0292

0.0314

± 3.203

± 1.953

± 0.488

0.0005

2.441

0.642

0.5

–0.5

0.0722

0.0742

0.0764

± 0.340

± 0.195

± 0.049

0.0010

0.244

0.064

0.05

–0.05

0.0722

0.0742

0.0764

± 0.054

± 0.063

± 0.006

0.0010

0.024

0.008

Note: Accuracies are valid for measurements following an internal E Series Calibration. Averaged numbers assume dithering and averaging
of 100 single-channel readings. Measurement accuracies are listed for operational temperatures within ± 1 °C of internal calibration
temperature and ± 10 °C of external or factory calibration temperature.

PCI.book Page 2 Wednesday, September 16, 1998 9:09 AM