Field Wiring 37

Chapter 2

Sense SCPs and

Output SCPs

Some SCPs provide input signal conditioning (sense SCPs such as filters
and amplifiers) while others provide stimulus to the measurement circuit
(output SCPs such as current sources and strain bridge completion).
In general, channels at output SCP positions are not used for external signal
sensing but are paired with channels of a sense SCP. Two points to remember
about mixing output and sense SCPs:

1. Paired SCPs (an output and a sense SCP) may reside in separate

VT1422As. SCP outputs are adjusted by *CAL? to be within a
specific limit. The Engineering Unit (EU) conversion used for a sense
channel will assume the calibrated value for the output channel.

2. Output SCPs, while providing stimulus to the measurement circuit,

reduce the number of external sense channels available to the VT1422A.

Figure 2-2 illustrates an example of "pairing" output SCP channels with sense
SCP channels (in this example, four-wire resistance measurements).

Note

Each channel line

represents both a

Hi and Lo input

SCP

3

(sense)

SCP

4

(output)

Input Mult

iplexer

Ch 24

Ch 31
Ch 32

Ch 39

Terminal Module

sense Hi

sense Lo

outpu

t Hi

output Lo

Figure 2-2. Pairing Output and Sense SCP Channels