handylab pH 12

16

ba75368e04

01/2011

4.2

Measuring

4.2.1

General information

You can measure the following variables:

z pH value
z ORP voltage

Preparatory

activities

Perform the following preparatory activities when you want to measure:

Caution
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result. The RS232 inter-
face is not galvanically isolated.

Note
Incorrect calibration of pH electrodes leads to incorrect measured values.
Calibrate regularly before measuring.

Temperature

sensor

You can measure with or without a temperature sensor. If a temperature sen-
sor is connected, it is indicated on the display by TP.

Note
The measuring instrument automatically recognizes the type of the tempera-
ture sensor used. Therefore, you can connect temperature sensors of the
NTC30 or Pt1000 type.

1

Connect the electrode to the measuring instrument.

2

Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.

3

In conjunction with the electrode, check or calibrate the mea-
suring instument. How to calibrate is described in section 4.3.

4

Select the measuring mode with <M>.