Calibration

evaluation

After the calibration, the measuring instrument automatically evaluates the
current status of the electrode. The asymmetry and slope are evaluated sep-
arately. The worst evaluation appears on the display.

Preparatory

activities

Display

Asymmetry
[mV]

Slope
[mV/pH]

-15 ... +15

-60.5 ... -58

-20 ... +20

-58 ... -57

-25 ... +25

-61 ... -60.5
or
-57 ... -56

Clean the electrode according
to the electrode operating
manual

-30 ... +30

-62 ... -61
or
-56 ... -50

E3

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

< -30 or
> 30

< -62 or
> -50

1

Switch on the measuring instrument with <ON/OFF>.

2

Connect the pH electrode to the measuring instrument.

3

Keep the buffer solutions ready.

4

Adjust the temperature of the solutions and measure the cur-
rent temperature if the measurement is made without temper-
ature sensor.