Operation

handylab pH 12

36

ba75368e04

01/2011

4.4.2

Saving automatically

The save interval (Int 1) determines the chronological interval between auto-
matic save processes.
After the fixed interval has expired, the current data record is transmitted to
the storage and to the interface.

Setting the save

interval

The default setting for the save interval (Int 1) is OFF.
By this, the AutoStore function is switched off.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min, 10 min,
15 min, 30 min, 60 min):

1

Press the <RUN/ENTER> key and hold it down.

2

Press the <STO> key. Int 1 appears on the display.

3

Set the required interval between the saving procedures with
<

> <>.

4

Confirm with <RUN/ENTER>.
The number of free memory locations appears on the display.

5

As soon as all of the 800 storage locations are occupied,
AutoStore is terminated (Int 1 = OFF).
If there are not enough storage locations available for your
measurements:

– Output and backup the data storage (see page 38) and

– clear the memory (see page 43).