Xylem handylab pH 12 User Manual

Page 18

Advertising
background image

Operation

handylab pH 12

18

ba75368e04

01/2011

AutoRead

(drift control)

The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of the
measured values.

For identical measurement conditions, the following criteria apply:

Note
The current AutoRead measurement can be terminated at any time (accept-
ing the current value) by pressing <RUN/ENTER>.

Reproducibility

Response time

Better than 0.02

> 30 seconds

1

Call up the pH measuring mode with <M> and <

> <> .

2

Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).

3

Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.

4

If necessary, start the next AutoRead measurement with
<RUN/ENTER>.

5

To terminate the AutoRead function: Press the <AR> key.