D i s – Xylem handylab pH 12 User Manual
Page 49

handylab pH 12
Operation
49
ba75368e04
01/2011
Display during the
pH calibration
Calibration interval
Date and time
6
Select the required display during the pH calibration with
<
▲> <▼>.
mV: Display of the current electrode voltage
/pH: Display of the buffer nominal value.
7
Confirm with <RUN/ENTER>. Int 3 appears on the display.
8
Set the required interval in days with <
▲> <▼>.
9
Confirm with <RUN/ENTER>.
The date flashes on the display.
2 d
°C
TP
mV/pH
mol/L
Auto Store
RCL
LoBat
CalError
ARng AR
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
3
nt
pH
Time
AutoCal DIN
AutoCal TEC ConCal
I
d
10
Set the date of the current day with <
▲> <▼>.
S
1
M
m
%
mg/l
cm
S/
cm
mbar
mV/pH
D i s
Store
Auto
Ident
LoBat
Tref25
nLF
/
K
%
Sal
°
AR
TP
C
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
O
UpH
AutoCal DIN
AutoCal TEC
Time
Baud
No.
Year
4
0 9 0
Day.Month