7 reset, Reset – Xylem handylab pH 12 User Manual
Page 51
handylab pH 12
Operation
51
ba75368e04
01/2011
4.7
Reset
You can reset (initialize) the measurement parameters and the configuration
parameters separately from one another.
Measurement
parameters
The following measured parameters (pH InI) are reset to the default condi-
tion:
Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.
Configuration
parameters
The following configuration parameters (InI) are reset to the delivery status:
Measuring mode
pH
Asymmetry
0 mV
Slope -59.16
mV/pH
Calibration procedures
AutoCal TEC
Temperature, manual
25 °C
Resolution (pH display)
0.01 (low resolution)
Display during the
pH calibration
Buffer nominal value
Baud rate
4800
Interval 1
(automatic save)
OFF
Interval 2
(for data transmission)
OFF