7 reset, Reset – Xylem handylab pH 12 User Manual

Page 51

Advertising
background image

handylab pH 12

Operation

51

ba75368e04

01/2011

4.7

Reset

You can reset (initialize) the measurement parameters and the configuration
parameters separately from one another.

Measurement

parameters

The following measured parameters (pH InI) are reset to the default condi-
tion:

Note
The calibration data gets lost when the measuring parameters are reset.
Recalibrate after performing a reset.

Configuration

parameters

The following configuration parameters (InI) are reset to the delivery status:

Measuring mode

pH

Asymmetry

0 mV

Slope -59.16

mV/pH

Calibration procedures

AutoCal TEC

Temperature, manual

25 °C

Resolution (pH display)

0.01 (low resolution)

Display during the
pH calibration

Buffer nominal value

Baud rate

4800

Interval 1
(automatic save)

OFF

Interval 2
(for data transmission)

OFF