4 saving, 1 saving manually, Saving 4.4.1 – Xylem handylab pH 12 User Manual

Page 34: Saving manually

Advertising
background image

Operation

handylab pH 12

34

ba75368e04

01/2011

4.4

Saving

The measuring instrument has an internal data memory. It can store up to
800 data records.
A complete data record consists of:

z Storage location
z Date/time
z Measured value
z Temperature
z Temperature measuring procedure
z ID number

You can transmit measured values (data records) to the data storage in two
ways:

z Save manually
z Switch on AutoStore (Int 1).

4.4.1

Saving manually

You can transmit a measured value to the data storage as follows:

1

Press the <STO> key.
The current number (location number No.) of the next free stor-
age location appears under the current measured value on the
display.

2

Confirm with <RUN/ENTER>.
The display switches to entering the ID number.

Ident

°C

2 4

TP

mV/pH

mol/L

8

Auto

TempError

RCL

LoBat

AutoCal DIN

CalError

ARng AR

AutoCal TEC ConCal

Time

Baud

Day.Month

Year

1

U

%

°F

S

mg/L

% ppm

ISE

4

2

99

6

pH

Store

No.

Advertising