handylab pH 12

Operation

27

ba75368e04

01/2011

4.3.2

AutoCal DIN

For this procedure, use two or three different standard buffer solutions ac-
cording to DIN 19266 in ascending or descending order (type A, C, D or F
with pH values at 25 °C of: 1.679 / 4.006 / 6.865 / 9.180).

Note
The buffer solutions are automatically recognized by the measuring instru-
ment. Depending on the instrument setting (see section 4.6 C

ONFIGURATION

),

the instrument displays the relevant buffer nominal value or the current elec-
trode voltage in mV.

Note
Skip the steps 2, 7 and 14 if you use a temperature sensor.

1

Press the <CAL> key repeatedly until the display Cd1 and the
function display AutoCal DIN appear. The sensor symbol dis-
plays the evaluation of the last calibration (or no sensor symbol
in the delivery state or after the measurement parameter has
been reset).

2

If necessary, enter the temperature of the buffer solution with
<

> <>.

3

Immerse the pH electrode in the first buffer solution.

4

Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example: