handylab pH 12
What to do if...
57
ba75368e04
01/2011
Obviously incor-
rect
measured values
Instrument does
not react to key-
stroke
Display
to
Probe symbol
flashes
Message
StoFull
Cause
Remedy
pH electrode:
– pH electrode unsuitable
– Use suitable electrode
– Temperature difference
between buffer and test
sample too high
– Adjust temperature of buffers
or test samples
– Measurement procedure not
suitable
– Follow special procedure
Cause
Remedy
– Operating condition
undefined or EMC load
unallowed
– Processor reset:
Press the <CAL> and
<ON/OFF> keys at the same
time and release them again.
The software version is
displayed.
Cause
Remedy
– Time-out of the interface
– Check the instrument that is
connected
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– All memory locations are full
– Output data storage and clear
data storage