Xylem handylab pH 12 User Manual
Page 32
Operation
handylab pH 12
32
ba75368e04
01/2011
8
To continue the two-point calibration, thoroughly rinse the
electrode with deionized water.
9
If necessary, enter the temperature of the second buffer solu-
tion with <
▲> <▼>.
10
Immerse the pH electrode in the second buffer solution.
11
Press the <RUN/ENTER> key. The second pH value appears
on the display.
12
Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <
▲> <▼>.
13
When the measured value is stable, press the <RUN/ENTER>
key. The value of the slope (mV/pH) appears on the display.
The probe symbol shows the evaluation of the current calibra-
tion.
14
Press the <RUN/ENTER> key. The value of the asymmetry
(mV) appears on the display.
AutoCal DIN
AutoCal TEC
6
mV/pH
mol/L
8
Auto Store
RCL
LoBat
CalError
ARng AR
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
TempError
mg/L
% ppm
ISE
8
4
2
O
SL
°C
TP
pH
ConCal