background image

handylab pH 12

Operation

43

ba75368e04

01/2011

4.4.4

Clearing the memory

With this function, you can delete stored data records. 800 storage locations
will then be available again.

Note
The Clear memory function only appears when there are data records stored
in the memory. Otherwise, the measuring instrument automatically switches
to the measuring mode.

Proceed as follows to clear all data records:

Note
The calibration data remain stored and can be called up.

1

Switch off the measuring instrument.

2

Press the <STO> key and hold it down.

3

Press the <ON/OFF> key.
The display test appears briefly on the display.
Subsequently, Sto clr appears.

4

Confirm the clearing process with <RUN/ENTER>.
Pressing any other key prevents the clearing, the data records
will remain stored.