The temperature measurement is absolutely essential for a reproducible pH
measurement. If the measurement is made without a temperature sensor,
proceed as follows:
4.2.2
Measuring the pH value
Setting the
resolution
The measuring instrument shows the pH value with a resolution of 0.01 or
0.001. (Default setting is 0.01). To change the resolution, press the <M> key
while pressing the <RUN/ENTER> key.
1
Measure the current temperature of the test sample using a
thermometer.
2
Enter the temperature on the instrument: Set the temperature
value of the test sample using <
▲ > <▼>.
1
Perform the preparatory activities according to section 4.2.1.
2
Immerse the pH electrode in the test sample.
3
Press the <M> key until pH appears on the status display. The
pH value appears on the display.