Operation

handylab pH 12

26

ba75368e04

01/2011

Three-point

calibration

Note
You can prematurely terminate the three-point calibration by pressing <M>.
The values of the two-point calibration for slope and asymmetry will then re-
main stored.

12

Press the <RUN/ENTER> key
to continue the three-point calibration.
Ct3 appears on the display.

13

Thoroughly rinse the electrode with distilled water.

14

If necessary, enter the temperature of the third buffer solution
with <

> <>.

15

Immerse the pH electrode in the third buffer solution.

16

Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display.
When the measured value is stable, AR disappears.
The instrument displays the value of the slope (mV/pH) and the
evaluation of the calibration in calibration range II.

17

Press the <RUN/ENTER> key.
The instrument displays the value of the asymmetry in calibra-
tion range II. The three-point calibration is completed.

18

To return to the measuring mode: Press the <M> key.