3 setting the profibus address, 4 gsd file, Setting the profibus address -5 – YSI DIQ/S 182-PR System 182 User Manual
Page 115: Gsd file -5, Section 7.3)
System 182
PROFIBUS connection
7 - 5
ba76024e01
02/2012
7.3
Setting the PROFIBUS address
Proceeding:
Setting the PROFIBUS
address on the
DIQ/S 182-PR
Note
To accept all settings, you have to highlight the Save and quit menu
item at the lower end of the setting table and press
g
. If you exit the
setting table via
m
,
e
/Quit or the Quit menu item, all changes are
ignored.
7.4
GSD file
The GSD file contains all necessary information on the Profibus module
and is required by the configuration program of the Profibus master.
The GSD file for the IQ S
ENSOR
N
ET
is found on the CD-ROM provided
(file name: IQMC06D1.GSD). The current GSD file is also provided on
the Internet under
.
1
Open the SETTINGS menu with
s
.
2
Select and confirm the PROFIBUS config. menu item with
w
and
g
.
3
Edit the setting table as described in section 4.3.2.