2 maintenance condition, Maintenance condition -9, G m e / / m c e/m g c g – YSI DIQ/S 182-PR System 182 User Manual

Page 69

Advertising
background image

System 182

Operation

4 - 9

ba76024e01

02/2012

4.4.2

Maintenance condition

When an IQ sensor is calibrated, cleaned, serviced or repaired, the
maintenance condition for the relevant IQ sensor should always be
switched on.

In the maintenance condition

 apart from the measured value display, the system does not react to

the current measured value or the condition of the selected
IQ sensor

 IQ sensor errors do not prompt changes in the conditions of linked

outputs.

The following diagram gives you an overview of when an IQ sensor is
in the maintenance condition.

= Maintenance condition ON

g m e

/

/

m

c

e/m

g

c

g

Measured value

display

The measured value

stops flashing

Measured value

display

The measured

value flashes

Measured value

display

flashes

Clean

Calibration

Step 1

Calibration

Step n

End

Maintanance

condition ON

manually

Maintanance

condition OFF

manually

Sensor

cleaning

active

Sensor

cleaning

OFF

(automatic sequence)

= Maintenance condition OFF

e

Switch

Display message:

on the

maintenance

condition

Continue/Cancel

Advertising