System 182

Technical data

10 - 3

ba76024e01

02/2012

Electrical data

DIQ/S 182-PR/24V

(24 V AD/DC version)

Electrical connections

DIQ/S 182-PR

(line power version)

Terminal strip inside the enclosure:

Electrical connections

DIQ/S 182-PR/24V

(24 V AC/DC version)

Terminal strip inside the enclosure:

Relay

(3 x)

Supply

Nominal voltage:

24 V AC/DC ± 10 %
protective low voltage
SELV (Safety
Extra Low Voltage)

AC frequency:

50/60 Hz according to
DIN IEC 60038

Connection:

2 pin

Line cross-section of connections:

Europe: 1.5 ... 4.0 mm

2

USA:

AWG 14 ... 12

Fuse rating on the operator side: max. 16 A

Power consumption

max. approx. 12 W

SENSORNET 1

SENSORNET 2

RED

RED

SHIELD

SHIELD

GREEN

GREEN

X3

X6

X2

X5

X17

X9

X15

X13

X11

X1

X4

X16

X8

X14

X12

X10

100...

240V AC

MAINS

AUXILIARY

R3

R2

R1

L1

N

240V AC

2A AC

240V AC

2A AC

240V AC

2A AC

AUXILIARY

VOLTAGE

SENSORNET 1

RED

SHIELD

GREEN

X3

X2

X17

X9

X15

X13

X11

X1

X16

X8

X14

X12

X10

24V AC DC

INPUT

POWER

AUXILIARY

R3

R2

R1

240V AC

2A AC

240V AC

2A AC

240V AC

2A AC

AUXILIARY

VOLTAGE

SENSORNET 2

RED

SHIELD

GREEN

X6

X5

X4

Output

Galvanically separated

Max. switching voltage

240 VAC or 24 VDC

Max. switching current

2 A (AC and DC)

Installation
requirements

Fuse rating on the operator side: max. 2 A

Relay functions

Programmable as:

– Opener or closer

– Limit monitor

– Monitoring of the warning and error

signals of the DIQ/S 182-PR

– Proportional frequency output

– Proportional pulse width output