Installation

3 - 39

ba76024e01

02/2012

Connecting lines to the

terminal strip

Fig. 3-31 Inserting lines

Note
The complete assignment of the terminal strip is shown in section 3.12.

11

Open the enclosure of the Universal Transmitter.

12

Screw a cable gland (pos. 1 in Fig. 3-31) with the sealing ring
(pos. 2) into the enclosure below the respective connections.

13

Loosen the coupling ring (pos. 3).

14

Feed the line through the cable gland in the enclosure.

15

Connect the wires to the terminal strip. While doing so, pay
attention to the specifications on the label located under the
terminal strip.

16

Tighten the coupling ring (pos. 3).