2 diq/s 182-pr/24v (24 v version), Diq/s 182-pr/24v (24 v version) -33 – YSI DIQ/S 182-PR System 182 User Manual

Page 47

Advertising
background image

System 182

Installation

3 - 33

ba76024e01

02/2012

3.8.2

DIQ/S 182-PR/24V (24 V version)

Warning

If the 24 V AC/DC supply is incorrectly connected, it may
represent a danger to life from electric shock. Pay attention to the
following points during installation:
The DIQ/S 182-PR Universal Transmitter may only be

connected by a trained electrician.

The 24 V AC/DC supply must meet the specifications quoted on

the name plate and in chapter 10 T

ECHNICAL

DATA

(protective low

voltage SELV).

The connection of the DIQ/S 182-PR Universal Transmitter to

the power supply may only be carried out when it is not
carrying any voltage.

When installed in a building, a switch or power switch must be

provided as an interrupt facility for the System 182.
The interrupt facility must

– be installed in the vicinity of the DIQ/S 182-PR Universal

Transmitter, easily accessible by the user, and

– be labeled as the interrupt facility for the DIQ/S 182-PR

Universal Transmitter.

Note
Rechargeable battery systems should have a deep discharge
protection. The DIQ/S 182-PR/24V does not have any built-in deep
discharge protection.

Materials required

 Wire end sleeves, suitable for the 24 V AC/DC feed line, with

suitable crimping tool

 1 x screwed cable gland with sealing ring (scope of delivery of the

Universal Transmitter).

Tools

 Cable stripping knife

 Wire stripper

 Phillips screwdriver

 Small screwdriver.

Preparing the 24 V AC/

DC line

1

Cut off the cable to the required length.

2

Strip the cable insulation for approx. 45 mm.

3

Bare the wires 1 and 2 and fit them with wire end sleeves.

Advertising