3 output data, 4 input data, Output data -9 – YSI DIQ/S 182-PR System 182 User Manual

Page 119: Input data -9

Advertising
background image

System 182

PROFIBUS connection

7 - 9

ba76024e01

02/2012

7.5.3

Output data

(1 byte - from the PROFIBUS master to the DIQ/S 182-PR)

7.5.4

Input data

(16 bytes - from the IQ S

ENSOR

N

ET

to the PROFIBUS master)

Address

Information

Bit

Offset 0h

Sensor number (Sxx) in the IQ S

ENSOR

N

ET

(Int8)

Bit 7-0

Address

Information

Bit

Offset 0h

Sensor number (Sxx) (Int 8)

Bit 7-0

Offset 1h

Sensor status (Int8)

Bit 7-0

Offset 2h

Sensor model (Int16)

Bits 15-8

Offset 3h

Bit 7-0

Offset 4h

Status info (Int16)

Bits 15-8

Offset 5h

Bit 7-0

Offset 6h

Measuring mode (Int8)

Bit 7-0

Offset 7h

Measured value status (Int8)
Main measured value bits 7-4
Secondary measured value bits 3-0

Bit 7-0

Offset 8h

Main measured value (IEEE-754 floating point, 32-
bit)

Bits 31-24

Offset 9h

Bits 23-16

Offset Ah

Bits 15-8

Offset Bh

Bit 7-0

Offset Ch

Secondary measured value (IEEE-754 floating point,
32-bit)

Bits 31-24

Offset Dh

Bits 23-16

Offset Eh

Bits 15-8

Offset Fh

Bit 7-0

Advertising