6 profibus error diagnosis, 7 profibus error elimination, Profibus error diagnosis -11 – YSI DIQ/S 182-PR System 182 User Manual

Page 121: Profibus error elimination -11

Advertising
background image

System 182

PROFIBUS connection

7 - 11

ba76024e01

02/2012

the status of the main measured value, bits 3-0 encode the status of the
secondary measured value.

Other data

All other data always consist of only a single byte (Int8).

7.6

PROFIBUS error diagnosis

A simple diagnosis of the PROFIBUS communication is possible via
the measured value and status display (status online/offline, see
section 4.2).

7.7

PROFIBUS error elimination

Note
Here you will find causes and actions to take of errors concerning the
PROFIBUS communication only. General errors of the IQ S

ENSOR

N

ET

system are dealt with in chapter 9 W

HAT

TO

DO

IF

....

Data transmission

between the Profibus

master and

DIQ/S 182-PR is faulty

The PLC does not

receive any plausible

input data

Cause

Remedy

– Incorrect wiring of the

PROFIBUS cable on the
connector

– Check/change the connections

(see section 7.2)

– Use checklist according to

section 7.1

– Incorrect setting of the

address

– Check/change the setting of

the address (see section 7.3)

– Incorrect protocol

– Check the version of the GSD

file

– Adapt the protocol

– DIQ/S 182-PR defective

– Send DIQ/S 182-PR to YSI

Cause

Remedy

– Input data and output data

are not consistent

– When programming the PLC,

define input data and output
data as consistent over the
entire data length

– The data interpretation of the

PLC is not correct

– Observe the data alignment of

the PLC data representation. If
necessary, exchange the high-
order and low-order bytes word
by word