3 replacing system components, 1 replacing passive components, Replacing system components -2 9.3.1 – YSI DIQ/S 182-PR System 182 User Manual

Page 126: Replacing passive components -2

Advertising
background image

What to do if ...

System 182

9 - 2

ba76024e01

02/2012

The selected system

language was not

activated for a sensor

9.3

Replacing system components

9.3.1

Replacing passive components

Passive components include all components that the controller cannot
recognize.
These include:

 MIQ power supply modules

 DIQ/JB

 DIQ/CHV

 Cables (SNCIQ, SACIQ).

Warning

If the DIQ/S 182-PR Universal Transmitter is opened and the MIQ/
PS power supply module is operating, there is a danger to life due
to possible hazard of electric shock from line voltage. Before
opening the Universal Transmitter, switch off the system and, if
necessary, all external relay circuits that present a hazard of
electric shock and secure the system against being switched on
again.

Only ever replace components when the system is switched off.
Defective components are removed in the reverse order to the
installation (see chapter 3 I

NSTALLATION

).

Cause

Remedy

– A system language was selected

that is not available in the
sensor.
The English language was
activated instead of the selected
system language.

– Contact YSI,

a software update is
required for the respective
components