Test port input (trace noise) – Atec Agilent-E5071C User Manual

Page 21

Advertising
background image

21

Test port input (trace noise)

Table 20.

Option 230/235/240/245/260/265/280/285/430/435/440/

445/460/465/480/485

Description

Specifi cation

SPD

Trace noise magnitude

1

9 kHz to 30 kHz (IFBW = 3 kHz)

0.004 dB rms

0.001 dB rms

30 kHz to 100 kHz (IFBW = 3 kHz)

0.003 dB rms

0.001 dB rms

100 kHz to 10 MHz (IFBW = 3 kHz)

0.003 dB rms

0.0005 dB rms

10 MHz to 4.38 GHz (IFBW = 70 kHz)

0.004 dB rms

0.001 dB rms

4.38 GHz to 5 GHz (IFBW = 70 kHz)

0.006 dB rms

0.0012 dB rms

5 GHz to 6 GHz (IFBW = 70 kHz)

0.006 dB rms

0.0012 dB rms

6 GHz to 7 GHz (IFBW = 70 kHz)

0.006 dB rms

0.0012 dB rms

7 GHz to 8.5 GHz (IFBW = 70 kHz)
(at maximum output power level of sweep range)

0.006 dB rms

0.0012 dB rms

Trace noise phase

2

9 kHz to 30 kHz (IFBW = 3 kHz)

0.035 deg rms

30 kHz to 10 MHz (IFBW = 3 kHz)

0.020 deg rms

10 MHz to 4.38 GHz (IFBW = 70 kHz)

0.035 deg rms

4.38 GHz to 5 GHz (IFBW = 70 kHz)

0.050 deg rms

5 GHz to 6 GHz (IFBW = 70 kHz)

0.050 deg rms

6 GHz to 7 GHz (IFBW = 70 kHz)

0.050 deg rms

7 GHz to 8.5 GHz (IFBW = 70 kHz)
(at maximum output power level of sweep range)

0.050 deg rms

1. The specifi cation might not be met at the following frequencies: 333.333 kHz, 406.25 kHz,

857.143 kHz, 928.571 kHz, 1.3 MHz, 2.4 MHz and 4.333333 MHz.

2. The specifi cation might not be met at 5 MHz or 50 MHz.

Advertising