Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485

Table 62.

Corrected system performance with type-N 75 Ω device connectors,

85036E calibration kit

Network analyzer:

E5071C

calibration kit:

85036E (type-N 75 Ω)

50 to 75 Ω adapters:

11852B

calibration:

full

2-port

IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C
±5 °C with < 1 °C deviation from calibration temperature, isolation calibration performed

Typical

(dB)

Description

10 MHz to 3 GHz

Directivity

37

Source match

33

Load match

39

Reflection

tracking

±0.015

Transmission tracking

±0.019

10 MHz to 3 GHz

S11 = S22 = 0; Source power = –10 dBm

Transmission using 85036E and 11852B

Transmission coefficient (dB)

10

1

0.1

0.01

Mag (dB)

10 0 –10 –20 –30 –40 –50 –60 –70 –80 –90

10 MHz to 3 GHz

S11 = S22 = 0; Source power = –10 dBm

Transmission using 85036E and 11852B

Transmission coefficient (dB)

100

10

1

0.1

Phase (deg)

10 0 –10

–20

–30

–40

–50

–60

–70

–80

–90

10 MHz to 3 GHz

S21 = S12 = 0; Source power = –10 dBm

Reflection using 85036E and 11852B

Reflection coefficient (linear)

10

8

6

4

2

0

Phase (deg)

0 0.2 0.4

0.6 0.8 1

10 MHz to 3 GHz

S21 = S12 = 0; Source power = –10 dBm

Reflection using 85036E and 11852B

Reflection coefficient (linear)

0.1

0.08

0.06

0.04

0.02

0

Mag (linear)

0 0.2 0.4 0.6 0.8 1

Transmission uncertainty 10 MHz to 3 GHz (typical)

Reflection uncertainty 10 MHz to 3 GHz (typical)

Magnitude

Phase

Magnitude

Phase

Corrected System Performance for 75 Ω Measurements with 11852B
50 to 75 Ω Minimum-Loss Pads (Supplemental Information)