Corrected System Performance

The specifications in this section apply to measurements made with the
Agilent E5071C network analyzer under the following conditions:

No averaging applied to data

Environmental temperature of 23 °C (±5 °C) with less than 1 °C deviation

from the calibration temperature

Response and isolation calibration performed

System dynamic range

Table 1.

Option 230/235/240/245/260/265/280/285/430/435/440/445/

460/465/480/485

Description

Specification

SPD

System dynamic range

1, 2

9 kHz to 300 kHz

72 dB

300 kHz to 10 MHz

IF bandwidth = 3 kHz

82

dB

10 MHz to 6 GHz

98 dB

6 GHz to 8.5 GHz

92 dB

9 kHz to 300 kHz

97 dB

115 dB

300 kHz to 10 MHz

107 dB

115 dB

10 MHz to 6 GHz

IF bandwidth = 10 Hz

123 dB

130 dB

6 GHz to 7 GHz

117 dB

128 dB

7 GHz to 8 GHz

117 dB

126 dB

8 GHz to 8.5 GHz

117 dB

124 dB

Figure 1. System dynamic range (specification and actual measurement data example,
IF bandwidth 10 Hz)

80

90

100

110

120

130

140

150

160

1.E+05

1.E+09

2.E+09

3.E+09

4.E+09

5.E+09

6.E+09

7.E+09

8.E+09

Test frequency [Hz]

System dynamic range [dB]

E5071C specification

1. The test port dynamic range is calculated as the difference between the test port rms noise fl oor

and the source maximum output power. The effective dynamic range must take measurement

uncertainty and interfering signals into account.

2. The specifi cation might not be met at 5 MHz or 50 MHz.