Test port input (stability) – Atec Agilent-E5071C User Manual
Page 23
Advertising
23
Test port input (stability)
1
Table 22
. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485
Description
Specifi cation
Typical
Stability magnitude
9 kHz to 3 GHz
±0.005 dB/°C
3 GHz to 6 GHz
±0.01 dB/°C
6 GHz to 8.5 GHz
±0.04 dB/°C
Stability phase
9 kHz to 3 GHz
±0.1 deg/°C
3 GHz to 6 GHz
±0.2 deg/°C
6 GHz to 8.5 GHz
±0.8 deg/°C
Table 23.
Option 2D5/2K5/4D5/4K5
Description
Specifi cation
Typical
Stability magnitude
300 kHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 20 GHz
±0.005 dB/°C
±0.01 dB/°C
±0.04 dB/°C
Stability phase
300 kHz to 3 GHz
3 GHz to 6 GHz
6 GHz to 20 GHz
±0.1 deg/°C
±0.2 deg/°C
±0.8 deg/°C
1. Stability is defi ned as a ratio measurement at the test port.