Test port input (stability) – Atec Agilent-E5071C User Manual

Page 23

Advertising
background image

23

Test port input (stability)

1

Table 22

. Option 230/235/240/245/260/265/280/285/430/435/440/

445/460/465/480/485

Description

Specifi cation

Typical

Stability magnitude

9 kHz to 3 GHz

±0.005 dB/°C

3 GHz to 6 GHz

±0.01 dB/°C

6 GHz to 8.5 GHz

±0.04 dB/°C

Stability phase

9 kHz to 3 GHz

±0.1 deg/°C

3 GHz to 6 GHz

±0.2 deg/°C

6 GHz to 8.5 GHz

±0.8 deg/°C

Table 23.

Option 2D5/2K5/4D5/4K5

Description

Specifi cation

Typical

Stability magnitude

300 kHz to 3 GHz

3 GHz to 6 GHz

6 GHz to 20 GHz

±0.005 dB/°C

±0.01 dB/°C

±0.04 dB/°C

Stability phase

300 kHz to 3 GHz

3 GHz to 6 GHz

6 GHz to 20 GHz

±0.1 deg/°C

±0.2 deg/°C

±0.8 deg/°C

1. Stability is defi ned as a ratio measurement at the test port.