E5092a configurable multiport test set – Atec Agilent-E5071C User Manual

Page 52

Advertising
background image

52

E5092A Configurable multiport test set

The section provides test input/output performance without calibration by the E5071C.

Table 52.

Test set input/output performance

Description

Specifi cation

Typical

Frequency range

50 MHz to 20 GHz

Damage level

20 dBm, ±35 VDC

Table 53.

Option E5092A-020 port performance

Description

Specifi cation

Typical

Load match (selected port)

SPDT switch

1

50 MHz to 2 GHz
2 GHz to 4 GHz
4 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz

17 dB
11 dB
8 dB
7 dB
4 dB
4 dB

SP4T switch

2

50 MHz to 2 GHz
2 GHz to 3 GHz
3 GHz to 8 GHz
8 GHz to 10 GHz
10 GHz to 18 GHz
18 GHz to 20 GHz

17 dB
11 dB
8 dB
7 dB
4 dB
4 dB

Load match (unselected port)

SPDT switch

1

50 MHz to 3 GHz
3 GHz to 10 GHz
10 GHz to 16 GHz
16 GHz to 18 GHz
18 GHz to 20 GHz

17 dB
11 dB
8 dB
6 dB
4 dB

SP4T switch

2

50 MHz to 3 GHz
3 GHz to 10 GHz
10 GHz to 16 GHz
16 GHz to 18 GHz
18 GHz to 20 GHz

17 dB
11 dB
8 dB
6 dB
4 dB

1. SPDT: Single-pole-double-throw switches. Applies to SW5, SW6, SW7, SW8, SW9 and SW10 in the E5092A. (See Figure 20.)
2. SP4T: Single-pole-four-throw switches. Applies to SW1, SW2, SW3 and SW4 in the E5092A. (See Figure 20.)