Cycle time for measurement completion – Atec Agilent-E5071C User Manual
Page 41
41
Cycle time for measurement completion
1, 2
(ms)
Table 40.
Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Sweep mode: Stepped, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: ON
Number of Points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth
Uncorrected
4.4
7.7
12
29
2-port cal
8.2
15
22
56
4-port cal
16
29
45
126
Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected
4.8
9.2
15
41
2-port cal
9.0
18
28
80
4-port cal
18
35
56
163
Start 1 GHz, stop 1.2 GHz, 1 kHz IF bandwidth
Uncorrected
54
202
397
1563
2-port cal
107
401
792
3125
4-port cal
213
800
1581
6247
Start 100 kHz, stop 3 GHz, 500 kHz IF bandwidth
Uncorrected
6.7
13
18
48
2-port cal
13
24
36
94
4-port cal
26
47
70
190
Start 100 kHz, stop 3 GHz, 100 kHz IF bandwidth
Uncorrected
7.1
14
22
60
2-port cal
14
27
42
119
4-port cal
27
53
82
237
Start 100 kHz, stop 3 GHz, 1 kHz IF bandwidth
Uncorrected
56
206
404
1583
2-port cal
112
410
805
3163
4-port cal
222
818
1608
6324
1. Supplemental performance data.
2. Measured with fi rmware version A.09.30.