Test port input (dynamic accuracy)
1
Table 24
. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485
Accuracy of the test port input power reading is relative to –10 dBm reference input
power level.
Description
Specifi cation
Typical
Dynamic accuracy magnitude
10 dBm
±0.207 dB
–30 dBm
±0.045 dB
–100 dBm
±2.00 dB
–110 dBm
±3.0 dB
Dynamic accuracy phase
10 dBm
±5.03 deg
–30 dBm
±0.30 deg
–100 dBm
±15.0 deg
Specification
±3.0 dB (at –110 dBm, Ref.= –10 dBm, typical)
1. Dynamic accuracy is verifi ed with the following measurements:
• compression over frequency
• IF linearity at two frequencies (1 MHz and 1.195 GHz) using a reference level of –10 dBm for an
input power range of 0 to –100 dBm.