Test port input (dynamic accuracy) – Atec Agilent-E5071C User Manual

Page 24

Advertising
background image

24

Test port input (dynamic accuracy)

1

Table 24

. Option 230/235/240/245/260/265/280/285/430/435/440/

445/460/465/480/485

Accuracy of the test port input power reading is relative to –10 dBm reference input
power level.

Description

Specifi cation

Typical

Dynamic accuracy magnitude

10 dBm

±0.207 dB

–30 dBm

±0.045 dB

–100 dBm

±2.00 dB

–110 dBm

±3.0 dB

Dynamic accuracy phase

10 dBm

±5.03 deg

–30 dBm

±0.30 deg

–100 dBm

±15.0 deg

Specification

±3.0 dB (at –110 dBm, Ref.= –10 dBm, typical)

1. Dynamic accuracy is verifi ed with the following measurements:

• compression over frequency

• IF linearity at two frequencies (1 MHz and 1.195 GHz) using a reference level of –10 dBm for an

input power range of 0 to –100 dBm.

0.01

0.1

1

10

-100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Accuracy (dB)

Test port power (dBm)

Magnitude

Accuracy (dB)

Test port power (dBm)

Phase

0.1

1

10

100

-100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Advertising