Test port input (group delay)

1

(continued)

Table 27.

Option 2D5/2K5/4D5/4K5

Description

Specifi cation

Supplemental information

Aperture (selectable)

(frequency span)/(number
of points - 1)

Maximum aperture

25% of frequency span

Minimum delay

Limited to measuring no more
than 180° of phase change
within the minimum aperture.

Accuracy

See graph below, typical

The following graph shows group delay accuracy with 3.5 mm (male) connectors, full
2-port calibration and a 10 Hz IF bandwidth.

Calibration kit (85052D). Insertion loss is assumed to be < 2 dB.

In general, the following formula can be used to determine the accuracy, in seconds, of a
specifi c group delay measurement:

±phase accuracy (degrees) / [360 x aperture (Hz)]

1. Group delay is computed by measuring the phase change within a specified step (determined by

the frequency span and the number of points per sweep).

C al pow er = -10 dB m ; M eas pow er = -10 dB m ; E lectrical Length = 10 m

Frequency = 1 G H z