50

Cycle time (ms)

1, 2

vs. IF bandwidth

Table 50.

All options

Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: OFF, NOP = 201

Option 230/235/240/245/260/265/280/
285/430/435/440/445/460/465/480/485

Option 2D5/2K5/4D5/4K5

Frequency
[Hz]

IFBW
[Hz]

Cycle time
[ms]

Frequency
[Hz]

IFBW
[Hz]

Cycle time
[ms]

4 G

500000

5.0

10 G

500000

2.8

4 G

400000

5.0

10 G

400000

2.8

4 G

300000

4.9

10 G

300000

2.8

4 G

200000

4.9

10 G

200000

2.8

4 G

150000

4.9

10 G

150000

2.8

4 G

100000

5.0

10 G

100000

3.0

4 G

70000

5.7

10 G

70000

3.8

4 G

50000

6.9

10 G

50000

4.9

4 G

40000

7.8

10 G

40000

5.9

4 G

30000

9.4

10 G

30000

7.5

4 G

20000

13

10 G

20000

11

4 G

15000

16

10 G

15000

14

4 G

10000

23

10 G

10000

21

4 G

7000

31

10 G

7000

29

4 G

5000

42

10 G

5000

40

4 G

4000

52

10 G

4000

50

4 G

3000

68

10 G

3000

66

4 G

2000

100

10 G

2000

98

4 G

1500

133

10 G

1500

132

4 G

1000

198

10 G

1000

196

4 G

700

278

10 G

700

276

4 G

500

391

10 G

500

389

4 G

400

487

10 G

400

485

4 G

300

648

10 G

300

646

4 G

200

970

10 G

200

968

4 G

150

1291

10 G

150

1289

4 G

100

1935

10 G

100

1933

4 G

70

2739

10 G

70

2739

4 G

50

3864

10 G

50

3869

4 G

40

4829

10 G

40

4837

4 G

30

6437

10 G

30

6449

4 G

20

9653

10 G

20

9671

4 G

15

12870

10 G

15

12892

4 G

10

19302

10 G

10

19330

1. Supplemental performance data.
2. Measured with fi rmware version A.09.30.