Cycle time for measurement completion – Atec Agilent-E5071C User Manual

Page 43

Advertising
background image

43

Cycle time for measurement completion

1, 2

(ms)

Table 42.

Option 2D5/2K5/4D5/4K5

Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: OFF

Number of points

51

201

401

1601

Start 300 kHz, stop 20 GHz, 500 kHz IF bandwidth

Uncorrected

23

36

39

43

2-port cal

45

71

77

85

4-port cal

88

142

155

169

Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth

Uncorrected

23

37

40

43

2-port cal

45

72

78

86

4-port cal

89

144

156

171

Start 300 kHz, stop 20 GHz, 1 kHz IF bandwidth

Uncorrected

60

212

410

1591

2-port cal

120

421

817

3180

4-port cal

237

840

1632

6358

Start 8 GHz, stop 18 GHz, 500 kHz IF bandwidth

Uncorrected

17

22

22

23

2-port cal

34

42

44

45

4-port cal

67

83

86

102

Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth

Uncorrected

18

22

23

24

2-port cal

34

42

44

47

4-port cal

67

84

87

106

Start 8 GHz, stop 18 GHz, 1 kHz IF bandwidth

Uncorrected

58

207

405

1583

2-port cal

115

413

807

3163

4-port cal

229

823

1612

6324

1. Supplemental performance data.
2. Measured with fi rmware version A.09.30.