Cycle time for measurement completion

1, 2

(ms)

Table 46.

Option 2D5/2K5/4D5/4K5

Sweep mode: Stepped, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: ON

Number of points

51

201

401

1601

Start 300 kHz, stop 20 GHz, 500 kHz IF bandwidth

Uncorrected

11

18

24

56

2-port cal

21

35

48

112

4-port cal

41

69

94

222

Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth

Uncorrected

11

20

28

69

2-port cal

22

38

54

137

4-port cal

42

75

107

272

Start 300 kHz, stop 20 GHz, 1 kHz IF bandwidth

Uncorrected

60

212

410

1591

2-port cal

120

421

817

3180

4-port cal

237

840

1632

6358

Start 8 GHz, stop 18 GHz, 500 kHz IF bandwidth

Uncorrected

8.4

14

19

48

2-port cal

17

27

38

94

4-port cal

32

52

74

188

Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth

Uncorrected

8.8

15

22

60

2-port cal

17

30

44

120

4-port cal

34

58

86

237

Start 8 GHz, stop 18 GHz, 1 kHz IF bandwidth

Uncorrected

58

207

405

1583

2-port cal

115

413

807

3163

4-port cal

229

823

1612

6324

1. Supplemental performance data.
2. Measured with fi rmware version A.09.30.