Cycle time for measurement completion – Atec Agilent-E5071C User Manual
Page 47
47
Cycle time for measurement completion
1, 2
(ms)
Table 46.
Option 2D5/2K5/4D5/4K5
Sweep mode: Stepped, analyzer display turned off with: DISP:ENAB OFF,
number of traces = 1, system error correction: ON
Number of points
51
201
401
1601
Start 300 kHz, stop 20 GHz, 500 kHz IF bandwidth
Uncorrected
11
18
24
56
2-port cal
21
35
48
112
4-port cal
41
69
94
222
Start 300 kHz, stop 20 GHz, 100 kHz IF bandwidth
Uncorrected
11
20
28
69
2-port cal
22
38
54
137
4-port cal
42
75
107
272
Start 300 kHz, stop 20 GHz, 1 kHz IF bandwidth
Uncorrected
60
212
410
1591
2-port cal
120
421
817
3180
4-port cal
237
840
1632
6358
Start 8 GHz, stop 18 GHz, 500 kHz IF bandwidth
Uncorrected
8.4
14
19
48
2-port cal
17
27
38
94
4-port cal
32
52
74
188
Start 8 GHz, stop 18 GHz, 100 kHz IF bandwidth
Uncorrected
8.8
15
22
60
2-port cal
17
30
44
120
4-port cal
34
58
86
237
Start 8 GHz, stop 18 GHz, 1 kHz IF bandwidth
Uncorrected
58
207
405
1583
2-port cal
115
413
807
3163
4-port cal
229
823
1612
6324
1. Supplemental performance data.
2. Measured with fi rmware version A.09.30.