System dynamic range – Atec Agilent-E5071C User Manual

Page 6

Advertising
background image

6

System dynamic range

(continued)

Table 2.

Option 2D5/2K5/4D5/4K5

Description

Specifi cation

SPD

System dynamic range

1, 2

300 kHz to 1 MHz

IF bandwidth = 3 kHz

70 dB

1 MHz to 10 MHz

82 dB

10 MHz to 100 MHz

95 dB

100 MHz to 6 GHz

98 dB

6 GHz to 8.5 GHz

92 dB

8.5 GHz to 10.5 GHz

80 dB

10.5 GHz to 15 GHz

75 dB

15 GHz to 20 GHz

71 dB

300 kHz to 1 MHz

IF bandwidth = 10 Hz

95 dB

105 dB

1 MHz to 10 MHz

107 dB

115 dB

10 MHz to 100 MHz

120 dB

129 dB

100 MHz to 6 GHz

123 dB

130 dB

6 GHz to 8 GHz

117 dB

129 dB

8 GHz to 8.5 GHz

117 dB

127 dB

8.5 GHz to 10.5 GHz

105 dB

115 dB

10.5 GHz to 15 GHz

100 dB

111 dB

15 GHz to 20 GHz

96 dB

105 dB

Figure 2. System dynamic range (specification and actual measurement data example,

IF bandwidth 10 Hz)


1. The test port dynamic range is calculated as the difference between the test port’s rms noise

fl oor and the source’s maximum output power. Effective dynamic range must take measurement

uncertainty and interfering signals into account.

2. The specifi cation might not be met at 5 MHz or 50 MHz.

Frequency [Hz]

S

ys

te

m

dy

n

a

m

ic

ra

n

g

e

[d

B

]

E5071C-2D5/2K5/4D5/4K5 specification

0.0E+00

5.0E+09

1.0E+10

1.5E+10

2.0E+10

80

90

100

110

120

130

140

150

160