8
Maximal is 6V. Channel two minimum value 0.004V, Maximal is 20V. }
<offset>:= { Default unit is "V". maximal value depends on the maximal
value depends on the version.}
<duty>:= {If wave type is square, range is from 20% to 80%. if wave type
is pulse, range is from 0.1% to 99.9%}
<symmetry> :={ 0% to 100%}
<phase>:= {0° to 360°}
<variance>:= Maximal is 2.222V, minimum value is 0.4mV.The default
unit is "V".
<mean>:= The range depends on Variance . The default unit is "V".
<delay>:= Maximal is Pulse Period, minimum value is 0.Unit is S.
QUERY SYNTAX
<channel>:BaSicWaVe?
<channel>:={C1, C2}
RESPONSE
<channel>:BSWV
<type>,<frequency>,<amplifier>,<offset>,<duty>,<symmetry>, <phase>
EXAMPLE 1
change channel one current wave type to ramp.
C1:BSWV WVTP,RAMP
EXAMPLE 2
Changes current signal frequency of channel one to 2000 Hz.
C1: BSWV FRQ, 2000HZ
EXAMPLE 3
set current signal amplifier of channel one.
C1: BSWV AMP, 3V
EXAMPLE 4
reads channel basic wave parameters from device.
C1:BSWV?
Return:
C1: BSWV WVTP,SINE,FRQ,1000,AMP,3,OFST,3,PHSE,0
RELATED COMMANDS
ARWV, BTWV, CFG, CPL, MDWV, SWWV
1.6. Modulate Wave Command
DESCRIPTION Set or get modulated wave parameters.
COMMAND SYNTAX
<channel>:MoDulateWaVe <parameter>