Get arbitrary wave data command – B&K Precision 4052-4055 - Programming Manual User Manual
Page 29

25
1.19. Get arbitrary wave data command
DESCRIPTION
This command changes the user defined memory unit arbitrary wave
data.
COMMAND SYNTAX
WaVe_DaTa <address>,<parameter>
<address>:= {M50~M59}
<parameter>:= {a parameter from the table below. }
Parameters
Value
Description
WVNM
<wavename> arbitrary wave name
TYPE
<type>
Arbitrary wave type .Note the value has
to be set to 5.
LENGTH
<length>
Arbitrary wave data Length. It must be
set to "32KB"
FREQ
<frequency> Arbitrary wave frequency.
AMPL
<amplifier>
Value of amplify.
OFST
<offset>
Value of offset.
PHASE
<phase>
Value of phase.
WAVEDATA
<wavedata>
Wave data.
Not: All parameters must be set in one command. If not, the command will not execute
successfully.
QUERY SYNTAX
WaVe_DaTa
RESPONSE FORMAT
WaVe_DaTa <parameter>
EXAMPLE Read device memory saved arbitrary data.
WVDT M50?
return:
WVDT POS, M51, WVNM, WAVE02, LENGTH, 32KB, TYPE, 5, WAVEDATA,\x00?\x03\x00\x06
\x00\t\x00\f\x00\x0f \x00\x12\x00\x15\x00\x19\x00\x1c\x00\x1f \x00"\x00%\x00(\x00+\x
00/\x002\x005\x008\x00;\x00>\x00A\x00E\x00H\x00K\x00N\x00Q\x00T\x00W\x00[\x00^\
x00a\x00d\x00g\x00j\x00m\x00q\x00t\x00w\x00z\x00}\x00\x80\x00\x83\x00\x87\x00\x8a
\x00\x8d\x00\x90\x00\x93\x00\x96\x00\x99\x00\x9d\x00\xa0\x00\xa3\x00\xa6\x00\xa9\x
00\xac\x00\xaf \x00\xb3\x00\xb6\x00\xb9\x00\xbc\x00\xbf \x00\xc2\x00\xc5\x00\xc9\x00\
xcc\x00\xcf \x00\xd2\x00\xd5\x00\xd8\x00\xdb\x00\xdf \x00\xe2\x00\xe5\x00\xe8\x00\xeb
\x00\xee\x00\xf1\x00\xf4\x00\xf8\x00\xfb\x00\xfe\x00\x01\x01\x04\x01\a\x01\n