B&K Precision 4052-4055 - Programming Manual User Manual

Page 21

Advertising
background image

17

<offset>:={ Default unit is "V". the maximal value depends on the
version.}
<duty>:={ If wave type is Square, range is from 20% to 80%
If wave type is pulse, range is from 0.1% to 99.9%}
<symmetry>:={ 0% to 100%}
<carr delay>:= {Maximal is Pulse Period, minimum valve is 0. Unit
is S.}
<variance>:={0.4mv to 666.6mv}
<mean>:={ the maximal value depends on the variance}



QUERY SYNTAX

<channel>:BursTWaVe? <parameter>

<parameter>:=<period>……


RESPONSE FORMAT

<channel>:BursTWaVe <type>|<state>|<period>……


EXAMPLE 1 Set channel one burst wave period to 1S.

C1:BTWV PRD, 1S


EXAMPLE 2

Set channel one burst wave delay to 0S

C1:BTWV DLAY, 0S


EXAMPLE 3

Reads channel two burst wave parameters that STATE is ON.

C2: BTWV?

Return:
C2:BTWV
STATE,ON,PRD,0.01S,STPS,0,TRSR,INT,TRMD,OFF,TIME,1,DLAY,2.4e-07S,,
GATE_NCYC,NCYC,CARR,WVTP,SINE,FRQ,1000HZ,AMP,4V,OFST,0V,PHSE,0

EXAMPLE 4

Reads channel two modulate wave parameters that STATE is OFF.

C2: BTWV?

Return:
C2: BTWV STATE,OFF