Use and Maintenance Manual Rev. 12-2012

ENGLISH

Resets the spindle reset values.

Displays the spindle reset values saved

Saves the spindle reset values.

Confirm

Goes back to the measurement screen

Return to main screen

8.3.4 Taper calibration

Saves the calibration

Confirm

Goes back to the reserved setup screen

Return to main screen

1. Open the spindle and fit the calibration tool fastening it to the adapter with the screw and nut provided.

2. Position the tool in such a way that the laser crosses its central part during normal reading movement.

3. Close the guard and press the button

4. When prompted on the screen, remove the tool and then fit it on the other side.

d

o

NoT

opeN

The

spINdle

.

5. When reading is complete, remove the calibration tool.

d

ANGeR

! d

o

NoT

peRfoRm

A

spIN

WITh

The

Tool

fITTed

!

6. Save the calibration.

7. Press

to go back to the main screen.

Reserved Setup