Brooks – Brooks Instrument GF81 User Manual

Page 41

Advertising
background image

4-3

Brooks

®

GF40/GF80/GF81 Devices

Installation and Operation Manual
X-TMF-GF40-GF80-Series-MFC-eng
Part Number: 541B161AAG
June, 2013

Section 4 Maintenance & Troubleshooting

GF40/GF80/GF81 Series
Performance

Possible Causes

Inaccurate flow.

Temperature shift (steady state or transient).

Inlet pressure shift (steady state or transient).

Power supply problem.

Electrical interference

Dirty gas chamber

Changes in gas.

Control problems. Can

Differential pressure not within operating range

not reach setpoint.

Inlet pressure not stable

Oscillation.

Zeroing problems, Indicated

Temperature shift (steady state or transient).

zero is not stable.

Inlet pressure shift (steady state or transient).

Power supply problem.

Electrical interference

Table 4-1 Environmental Factors

4-4 Troubleshooting Checklist

1. Check environmental factors that could affect changes to GF40/GF80/

GF81 Series performance. The most common environmental factors are
listed in Table 4-1.

2. Check supply voltage and check for a consistent ground.

3. Insure OEM tool setpoint matches the setpoint at the GF40/GF80/GF81

Series. Observe for consistency.

4. Verify isolations valves are open and the gas supply is turned on. Then

verify operating pressures are within operating ranges.

5. Check GF40/GF80/GF81 Series voltage response by moving the

setpoint back and forth.

Observe for voltage changes.