Section 12

Default Configuration

12.1 User configuration table

User Configuration

Option

Default Value

Possible Values

Selected Values

Date

MEASUREMENT

pH

11.20

09.00 - 12.50

Tgaz/Twater %

20

0000 - 0250

On-line measure time

10 minutes

0 - 9999 minutes

Smart rinsing

No

Yes or No

Max. rinse time

45 minutes

0 - 500 minutes

Reactivation
frequency

24 hours

0 - 999 hours

Reactivation fixed date SMTWTFS 15:00:00

Any day and time

Graph time base

12 hours

4 hours
8 hours
12 hours
16 hours
20 hours
24 hours

ALARMS / ALARM 1

Mode

Limit

Limit
Active channel
Sample flow
None

Attrib

Channel 1

Channel 1

Limit

1.00 ppm

0 - 9990 ppm

Direction

Up

Up
Down

Delay

0 seconds

0 - 999 seconds

Hysteresis

0%

0 - 10%

Relay

N.O.

N.O.
N.C.

ALARMS / ALARM 2

Mode

Limit

Limit
Active channel
Sample flow
None

Attrib

Channel 1

Channel 1

Limit

1.00 ppm

0 - 9990 ppm

Direction

Up

Up
Down

Delay

0 seconds

0 - 999 seconds

Hysteresis

0%

0 - 10%

Relay

N.O.

N.O.
N.C.