5 analyzer stabilization, 6 analyzer setup, 5 analyzer stabilization 4.6 analyzer setup – Hach-Lange POLYMETRON 9245 User Manual User Manual
Page 54: Installation

52
Installation
4.5 Analyzer stabilization
At this stage the analyzer has been completely installed, but needs to run for a period of time to
stabilize.
1.
Press Start on the main menu to start the measurement process.
2.
Leave the system to run for a couple of hours before starting any calibrations.
4.6 Analyzer setup
After the analyzer has stabilized, the system and user parameters must be set, followed by an
initial calibration.
1.
SYSTEM SETUP
.
2.
USER SETUP
- see
3.
CALIBRATION
- see
and launch a two point calibration.
Note:
The complete calibration cycle will last around 1 hour. However, the instrument cannot be calibrated
until at least one complete measurement cycle has been successfully performed. An attempt to calibrate the
instrument before this will result in a “Not authorized” message being displayed.