3.8 Manual calibration cycle

The following section gives an overview of the manual calibration process. For details on setting
up and running this processes, refer to the section entitled

Calibration on page 85

.

This process calculates the slope and offset of the ISE sodium electrode and the reference
electrode. It is based on the measurement of potential and temperature of two different samples
of known sodium concentration.

The concentration values of the two calibration solutions must be entered into the analyzer prior
to calibration (see

Two point calibration on page 89

). The lower concentration should be greater

than 50 ppb and the relationship between low and high concentration should be 10 or more. The
recommended low value is 100 ppb and the high value is 1,000 ppb.

The cycle is as follows:

1.

Phase 1:

The electrode is reactivated. This consists of injecting Sodium Nitrate into the

solution to maintain a high level of sodium measurement for a minimum of 5 minutes.

2.

Phase 2:

The measurement cell is rinsed.

3.

Phase 3:

The overflow vessel is rinsed and refilled with fresh sample.

4.

Phase 4:

The overflow vessel is drained, and the system waits for the operator to fill the

overflow vessel with the low value calibration solution.

Figure 8 Manual calibration cycle