12 fill electrolyte reservoir – Hach-Lange POLYMETRON 9245 User Manual User Manual

Page 53

Advertising
background image

51

Installation

4.4.12 Fill electrolyte reservoir

1.

The electrolyte reservoir is
located at the back of the
analyzer.

2.

Take the KCl electrolyte
bottle and insert the tip of
the tapered spout into the
reservoir inlet tube
(illustrated right) as far as
it will go but without
exerting any extra
pressure.

3.

Squeeze on the bottle as
many times as necessary
to fill the reservoir to
about 3/4 of its capacity. If
you have any difficulty
filling the reservoir, raise
the spout of the bottle
very slightly to avoid an
air lock.

4.

Using thumb and
forefinger, pump on the
electrolyte tube between
the reservoir and the
reference electrode to
remove any air bubbles
that may have formed.

5.

If necessary, clean any
KCl drops from the
analyzer and the
reservoir.